Table 9
Module
Pump
Autosampler
TCC
Detector
Agilent 1290 Infinity LC System Manual and Quick Reference
Data Acquisition in Method and Run Control View
Method parameters for first separation run
Parameter
Solvent A
Solvent B
Flow rate
Initial Composition
Gradient Timetable
Stop Time
Injection
Needle wash
Column
Temperature
Signal A
Peak width
Store Spectra
Quick Start Guide
Setting
Water
Acetonitrile
0.4 ml/min
60 % A, 40 % B
At 4 minutes 20 % A, 80 % B
5 minutes
1 µl
Flush port, 6 s
ZORBAX Eclipse Plus C18
1.8 µm, 2.1 mm x 50 mm i.d.
40 °C
250 nm, bw 100 nm, ref
360 nm, bw 100 nm
0.025 min (10 Hz)
All
5
97