Optimization of the Agilent 1290 Infinity LC System

How to Achieve Lowest Carry Over

How to Achieve Lowest Carry Over

Carryover is measured when residual peaks from a previous active-containing
injection appear in a subsequent blank solvent injection. There will be carry
over between active injections which may lead to erroneous results. The level
of carryover is reported as the area of the peak in the blank solution expressed
as a percentage of the area in the previous active injection. The Agilent 1290
Infinity autosampler is optimized for lowest carryover by careful design of the
flow path and use of materials in which sample adsorption is minimized. A
carryover figure of 0.002 % should be achievable even when a triple
quadrupole mass spectrometer is the detector. Operating settings of the
autosampler allow the user to set appropriate parameters to minimize
carryover in any application involving compounds liable to stick in the system.
The following functions of the autosampler can be used to minimize carryover:
• Internal needle wash
• External needle wash
• Needle seat backflush
• Injection valve cleaning
The flow path, including the inside of the needle, is continuously flushed in
normal operation, providing good elimination of carryover for most situations.
Automated delay volume reduction (ADVR) will reduce the delay volume but
will also reduce the flushing of the autosampler and should not be used with
analytes where carryover might be a problem.
The outside of the needle can be washed using a wash vial in a specific
location or the needle can be washed using the flush port. If a wash vial in a
tray location specified by the user is chosen then this vial should have no
septum and should contain a solvent suitable for washing the sample from the
needle. The septum is not used to avoid wiping contamination off the needle
on the downstream only to re-apply it on the upstroke. The needle can be
dipped into the vial multiple times. This will be effective in removing a small
degree of carryover but for more effective washing of the outside of the needle
use the flushport.
66
Agilent 1290 Infinity LC System Manual and Quick Reference