MAS2600: Maintenance and Troubleshooting
August 2023

MAS 2600 Test Sheet

Name
Ship name
MAS 2600 Transducer serial no.
Tank name
Liquid in tank
The tank is
Step
1
Measure voltage between terminal 5
and 6.
2
Measure nipple voltage between
terminal 5 and 6.
3
Measure sensor excitation voltage
between terminal 7 and 10.
4
Measure sensor excitation voltage
between terminal 8 and 9.
5
Measure current in supply lead to
terminal 5.
6
Measure current in supply lead to
terminal 6.
7
Measure sensor resistance with sensor
cable unconnected between:
Terminal 7 to cable shield.
Terminal 8 to cable shield.
Terminal 9 to cable shield.
Terminal 10 to cable shield.
Yard
MAS 2600 Amplifier serial no.
HFO
DO
Full
Empty
Ballast
Other
Result
Unit
V DC
V AC
V DC
mV DC
mA DC
mA DC
M Ohm
M Ohm
M Ohm
M Ohm
Reference Manual
RM 6000-01-001-24
ATT.
Emerson
Marine Solutions
Damcos A/S, Service Department
FAX.: +45 5578 7272
NB no.
Other
Limits
Voltage should be in the range 17 to 22 V
DC. Check power supply.
Voltage must not exceed 2 V AC.
Check power supply.
Voltage should be in the range 3.0 and
6.2 V DC.
Voltage should be in the range -15 to 100
mV DC.
Current should be in the range 4 to 20
mA and the same value as measured in
point 6. if not the amplifier may be
defect.
Current should be in the range 4 to 20
mA and the same value as measured in
point 5. if not the amplifier may be
defect.
If the measured resistance is less than 1
M Ohm the sensor is considered defect.
49