7.7.3
Stirring
Procedure
7.7.4
Heating and Adjusting Temperature
Procedure
22 January 2019
Installation and Initial Operation
INFORMATION
The exit gas cooler only works with activated temperature con-
trol (Parameter Temperature ON).
3.
Use your hands to check whether the exit gas cooler is begin-
ning to cool down. Open the control valve for water flow on the
basic unit, if necessary.
When the circuit is full, water escapes from the overflow on
the vessel jacket or cold finger and flows into the water outlet.
4.
Wait for min. 1 minute.
This ensures that no air is left in the temperature control cir-
cuit.
Bioreactor is running with temperature switched on
To test the stirrer, proceed as follows:
1.
On the operating panel for the parameter Stirrer, set a low set-
point.
For further information about the different ranges of rotation
speed refer to the specification in main chapter "Technical
Data".
2.
Switch the parameter on.
For the rest of the procedure, allow the bioreactor to run with the
temperature switched on and the stirrer running.
Bioreactor is running with temperature and stirrer switched on
To test the heating and adjust the temperature, proceed as follows:
1.
On the operating panel, set a high setpoint for parameter
Temperature, e.g. 45 °C..
The water supply for cooling is stopped; the system heats up.
a)
Temperature control system with electrical heater pad
and cold finger:
The heater pad starts to heat up.
b)
Water temperature control system:
Labfors 5 – Version for Microorganisms
Page 103 of 205