2 EXP. COMPARISON FUNCTIONS
- Hold value B HI Limit (HI-B) & Difference value-HI Limit (DF-HI)
- Hold value B LO Limit (LO-B) & Difference value-LO Limit (DF-LO)
Chapter
- OK
2
* The Hysteresis setting is not reflected in the Difference-HI/LO Limit comparison operation.
* If the Comparison Timing condition is not met, the comparison result of the value targeted for the
Comparison Timing turns OFF.
* OK is displayed each on the LCD screen as hold value A is OK, and hold B and the difference
value are OK.
2-3.
Relative Value Comparison
ON / OFF
How to set
SET.
On two hold points (A and B) in Double hold, HI/LO Limit comparisons of the comparison
standard hold value, and HI/LO Limit judgments of the difference (relative value) between the
comparison standard hold value and the other hold value can be made.
120
ON condition: Hold value B > HI-B Limit, or Difference value > Difference value-HI Limit
OFF condition: Hold value B ≦ HI-B Limit, and Difference value ≦ Difference value-HI Limit
ON condition: Hold value B < LO-B Limit, or Difference value < Difference value-LO Limit
OFF condition: Hold value B ≧ LO-B Limit, and Difference value ≧ Difference value-LO Limit
ON condition: All of HI-A, LO-A, HI-B, LO-B, DF-HI, and DF-LO are OFF.
OFF condition: Any of HI-A, LO-A, HI-B, LO-B, DF-HI, and DF-LO is ON.
- External I/O pin assignments for Comparison Standard B
A1
-
A2
OUT
HI-B & DF-HI
A3
OUT
A4
OUT
A5
OUT
A6
OUT
LO-B & DF-LO
A7
-
A8
OUT
HOLD END(H/E)
A9
OUT
A10
OUT
A11
OUT
A12
OUT
External I/O
FCN-365P024-AU/FCN-360C024-B
Initial value: OFF
Exp. Setting
→
→
COM1
COM2
WORK0
HI-A
WORK1
OK
WORK2
LO-A
WORK3(LOCK)
T/H
COM1
COM2
SECTION
RUN
D/Z
EVENT
GRAPH TRIG
SIF
CAL0
SIF
CAL1
(in expansion double hold only)
Exp. Comp. Set.
→
B1
-
IN
B2
IN
B3
IN
B4
IN
B5
IN
B6
B7
-
IN
B8
IN
B9
IN
B10
IN
B11
IN
B12
Page 1