2.6.4 Networking Overview

CLARUS® 500 can run as a standalone instrument storing patient
and exam data locally, or it can be connected to a network server
or EMR system to share data and worklists in a central location.
Adding peripheral equipment
may result in noncompliance with the safety requirements of IEC
60601-1.
You are responsible for ensuring that the system meets the
u
safety requirements of IEC60601-1.
Place any AC-powered, non-medical device peripherals at least
u
1.5 m away from the device and connect them to a separation
device, or else use an isolation transformer.
Connection to It-Networks including other equipment
could result in previously unidentified risks related to patients,
operators or third parties.
Changes to the IT networks could introduce new risks that require
additional analysis.
Changes to the IT networks include:
Changes to network configurations
u
Connections of additional items
u
Disconnection of items
u
Update of equipment
u
Upgrade of equipment
u
ð The Responsible Organization should identify, analyze, evaluate
and control these risks.