Test
Band (MHz)
Frequency
(MHz)
385
380 - 390
450
430 - 470
710
704 - 787
745
780
810
800 - 960
870
930
1720
1700 - 1990
1845
1970
2450
2400 - 2570
5240
5100 - 5800
5500
5785
Table 3: Test specifications for enclosure port immunity to RF wireless communications equipment.
20 / 192
1
UT is the a.c. mains voltage prior to application of the test level.
1.7 Wireless Communications
Service
Modulation
TERTRA 400
Pulse 18 Hz
GMRS 460,
FM ± 5 kHz
FRS 460
deviation
1 kHz sine
LTE Band 13,
Pulse 217 Hz
17
GSM
Pulse 18 Hz
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
Pulse 217 Hz
CDMA 1900;
GSM 1900;
DECT; LTE
Band
1,3,4,25;
UMTS
Bluetooth,
Pulse 217 Hz
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
WLAN
Pulse 217 Hz
802.11 a/n
Maximum
Distance (m) Immunity
Power (W)
1.8
0.3
2
0.3
0.2
0.3
2
0.3
2
0.3
2
0.3
0.2
0.3
2660021171806 Rev. A 2019-01
Instructions for Use
CLARUS® 500
Test Level (V/
m)
27
28
9
28
28
28
9