Instructions for Use
CLARUS® 500
WARNING!
WARNING!
WARNING!
WARNING!
WARNING!
2660021171806 Rev. A 2019-01
Modifying this device
may increase risks and decrease the service life of the device.
Do not use in the presence of flammable gasses.
u
Do not use in oxygen rich environments.
u
Using this device in the presence of flammable gases
may cause ignition and result in a fire hazard.
Do not use this device in the presence of flammable anesthetics
u
or oxidizers such as pure oxygen or nitrous oxide.
Use of this equipment adjacent to or stacked with other
equipment
should be avoided because it could result in improper operation.
If such use is necessary, this equipment and the other
u
equipment should be observed to verify that they are operating
normally.
Use of accessories, transducers, and cables other than those
specified or provided by the manufacturer of this equipment
could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper
operation.
Use only the accessories, transducers, and cables specified or
u
provided by ZEISS.
Adding peripheral equipment or replacing parts
may result in noncompliance with the safety requirements of IEC
60601-1.
You are responsible for ensuring that the system meets the
u
safety requirements of IEC 60601-1.
Only approved equipment to their respective IEC/ISO standards
u
such as IEC 60601-1 or IEC 60950 are allowed to be connected
to the Signal Input/Output parts (SIP/SOPs).
Place any AC-powered, non-medical device peripherals at least
u
1.5 m away from the instrument and connect them to a
separation device, unless an isolation transformer is used.
1 Safety and Certifications
1.3 Warnings, Cautions, and Notes
13 / 192