Instructions for Use
CLARUS® 500
Prerequisite
Action
This callout image has two
screenshots-- one for USA;
another for the rest of the world.
2660021171806 Rev. A 2019-01
7.4.5 Capturing a Montage Image (2-6 Image Custom
Montage)
Figure 18: Custom Montage Example
When you capture a custom Montage image, CLARUS® 500 takes
a series of widefield images and stitches them together. You can
decided how many images to include in the montage. For each
image, you can choose any internal fixation target location or have
the patient focus on the external fixation target (Using the External
Fixation Target [} 123]).
To capture a Montage image:
þ The device is prepared. (See Preparing the Device [} 99].)
þ The patient is prepared. (See Preparing the Patient [} 100].)
1. Under Scans, select UWF.
2. Select the Scan Type (Color, IR, FAF-Green, or FAF-Blue). For
more information about Scan Types, refer to: Scan Types
[} 108].
3. Under Settings, select Myd if the patient's pupil is dilated and
Non Myd if the patient's pupil is not dilated.
4. Using the touchscreen or mouse, select a fixation target (1) in
the Preview panel.
5. Instruct the patient to look inside the imaging aperture, focus
on the green fixation target and blink naturally.
7.4 Acquire Images
1
7 Operation
117 / 192