Prerequisite
This callout image has two
screenshots-- one for USA;
another for the rest of the world.
Action
Result
Action
2660021171806 Rev. A 2019-01
To manually position the internal fixation target:
þ A patient is being prepared for Scan Acquisition and you
want to change or set the internal fixation target.
1. In the IR Preview panel, double-click (or press the touchscreen
and hold) to select a position for the internal fixation target (1).
ü The internal fixation target moves to your selected location
in both the Preview panel and the patient's FOV.

7.4.11 Using the External Fixation Target

The external fixation target is a light you can position to guide the
patient's gaze. The external fixation target is in the accessory kit.
Use this target when the patient cannot see the internal fixation
target.
To install the external fixation target:
1. Remove the external fixation target from the accessory kit.
2. Install the external fixation target (1) onto mount (2) on the
patient support (3).
1
7 Operation
7.4 Acquire Images
123 / 192