5 Configuration
5.11 System Administration
User Name
Zeiss
ZeissAdmin
Technical Support
Table 8: Initial User Names and Passwords
NOTE
5.11.3.1.1 Internal Network Ports
NOTE
5.11.3.1.2 External Network Ports
5.11.3.2.1 Required Internal Network Ports
5.11.3.2.2 Required External Network Ports
Service
Database inbound
82 / 192
Zeiss instruments initially have three user names and passwords.
Initial passwords are shown in the table below. Change these
passwords before using the instrument.
Type
Administrator
Administrator
Administrator

5.11.3 Networking

5.11.3.1 Network Controllers and IP Addressing
The CLARUS® 500 is not compatible with networks using
IPv4 addressing in the range of 192.168.111.0 to
192.168.111.254.
The CLARUS 500 contains two network controllers; Internal
Network and External Network.
Do not rename the Internal Network.
Internal Network is for instrument use only with the following
assignments:
• Static IPv4 address: 192.168.111.2
• Subnet mask: 255.255.255.0
The External Network is for CLARUS® 500 connectivity and
automatically picks up an IP address.
5.11.3.2 Required Network Ports
The network ports listed in this section are required for proper
operation of the CLARUS® 500.
• Internal Communication Port: TCP 80
• Internal Communication Port: TCP 22
Password
November171846
November171846
<unique>
Instructions for Use
CLARUS® 500
Purpose
Instrument User
Instrument Adminis-
tration
Zeiss Technical Support
TCP
UDP
18081
2660021171806 Rev. A 2019-01