2 Hardware Installation (Differences)
Verifying EI Scan System Performance
Verifying EI Scan System Performance
Verifying performance of the G2570A GC/MSD system involves creating and saving
the GC method (OFN_SN) in addition to the steps described in the Hardware
Installation Manual (Part No. G2589-90006).
The GC parameters for OFN_SN are found in Appendix D.
Preparing the MSD
1 Start up the MSD.
Verify that the foreline pump and front fan turn on and that the foreline pump
stops gurgling within 60 seconds.
2 From the MSD Productivity Software, start the instrument session.
3 Switch to the Instrument view and verify that DEFAULT.M is the loaded
method.
4 From the Instrument view, verify that both icons and pulldown menu items
for GC parameters in the Instrument view are NOT accessible.
40