2 Hardware Installation (Differences)
Installing the Column Inlet
6 Inspect the end for jagged edges or burrs.
If the break is not clean and even, repeat steps 4 and 5.
7 Wipe the outside of the inlet column end with a lint-free cloth moistened
with methanol.
8 Position the column so it extends 4 to 6 mm past the end of the ferrule.
9 Slide the septum so that it is even with the bottom of the column nut guide
tube.
m’¤d"›apm˜ ¤z
F=¤4amV˜¤›
"zadd"©˜ pd¤hm
–GQ¬˜ mdF›˜ pd¤hm˜¤›
Fz›¤h
10 Insert the column nut into the inlet without letting the ferrule slip off the
column and finger tighten the nut.
11 Adjust the column position so the septum is touching the bottom of the
column nut guide tube.
12 Tighten the column nut an additional 1/4 to 1/2 turn (you should hear two
squeaks).
The column should not slide with a gentle tug.
26
4 to 6 mm