2 Hardware Installation (Differences)
Installing the Column in the GC/MSD Interface
12 Tighten the interface nut an additional 1/4 to 1/2 turn with a 1/4 inch open-
end wrench.
Check the tightness after one or two heat cycles.
13 If available, install the G2613A Autoinjector accessory and connect its
cable to the top "Injector" port in the back of the GC. Also install the
injector mounting bracket and post (found in the GC Ship Kit) next to the
GC power switch.
"camV˜p’›
mbF4›p˜p¤m›amV˜ "4cF›
14 Turn the GC on.
15 Verify that the injection port temperature is set to OFF (see the Status/
Settings/Inject screen).