11 Sequence execution can be paused by pressing the PAUSE key (F6) in the
Status/Automation screen. If a run is in progress, the GC will complete the
run and then pause the sequence.
After the current run is completed, the MSD ChemStation will initiate the next line
in the Sample Log Table even if the GC sequence is paused. The ChemStation will
then wait for a start signal from the GC.
12 Sequence execution can be resumed by pressing the PAUSE key again.
As the next line in the MSD Sample Log Table has been initiated, there is no need
to select Position and Run from the MS Top view.
13 The paused sequence can be canceled by pressing the Stop Seq key (F7)
in the Status/Automation screen.
To cancel the MSD sequence, press ABORT in the MS TOP view.
14 An active sequence can be stopped by pressing the Stop Seq key (F7) in
the Status/Automation screen. If a run is in progress, the GC will complete
the run and then stop the sequence.
— A stopped sequence cannot be resumed.
— To cancel the MSD sequence, press ABORT in the MS Top view.
15 An active run and sequence can be stopped by pressing the Stop button on
the GC Keypad or the STOP key (F7) in the Status screen.
To cancel the MSD sequence, press ABORT in the MS Top view.
3 Operational Differences
Automatic injection
83