2 Hardware Installation (Differences)
Installation Checklist
o
Performance Verification
o MSD powered on
o MSD Instrument session started
o Verify that GC icon and parameters are NOT accessible
o Save DEFAULT.M
o Load and resave OFN_SN.M from the \5973N subdirectory
o MSD source and quad temperature zones heating
o OFN_SN GC method created and saved using the GC Control Module
o Correct injection mode (manual or autoinjector) set in Status/Automation
screen
o System pumped down for at least 60 minutes
o GC oven set to 150° C and the column flow to 1.0 mL/min
o Checkout Tune performed
o Tune Evaluation performed
o OFN_SN GC method reloaded without saving changes
o OFN sample (1 pg/uL) ready
o Sensitivity Check method selected from MSD ChemStation
o Ready to Inject status box appears in MSD ChemStation screen
o Sample injected manually or with autoinjector
o MSD start signal received from GC start
o MSD stop signal received when GC stops
o EI Sensitivity performance verified