3 Operational Differences

Manual injection

Manual injection
Manual injection can be performed with or without an autoinjector connected to
the GC. The autoinjector can be bypassed by:
• Selecting
• Placing the autoinjector onto the injector mounting bracket located at the right
The Status screen of the GC Control Module should now be titled
Important
Running a sample using the G2570A GC/MSD system requires that you start the
MSD acquisition process before starting the GC. The GC start signal must not be
initiated until the
ChemStation monitor screen. Selecting the Start Run button in this status box is
unnecessary and will result in incorrect retention times.
78
in the Sequence Type field of the Status/Automation screen in the
None
GC Control Module, and
side of the GC.
Acquisition - Waiting for remote start...
Manual Run Status
status box appears on the
.