9 Display the Diagnostics screen via the Status/Service/Diagnostics path.
Detector Test key is inactive when the MSD is the detector.
10 Choose the Inlet Test key (F3 key).
The 6850A GC is shipped from the factory with an inlet nut containing a blank
ferrule; the Leak Test and Split Vent Test is based on having a blank ferrule
installed in the inlet. If you are running these tests after using the system,
remember to re-install the blank ferrule before proceeding.
11 From the Inlet Test screen, perform the Leak Test (F8 key) and then
perform the Split Vent Test diagnostics (F7 key), following the
instructions for both tests.
Correct any inlet problems resulting from these tests before continuing.
12 Run the Keyboard Test (F5 key in the Diagnostics screen) after successful
completion of the Inlet Test diagnostics pass.
13 Turn off the power to the GC.
2 Hardware Installation (Differences)
Installing the GC
19