To acquire data from a single injection of a single sample
1 Load the autoinjector with wash and waste bottles placed in the correct
places in the turret.
2 Place the sample in position 1 in the turret.
3 Load the MSD method from the ChemStation and the GC method from the
Control Module.
4 Set the appropriate injector parameters (accessed from the Status/
Automation screen).
5 In the Status/Automation screen, set the following sequence parameters:
— First Vial to 1
— Last Vial to 1
— # Inj per Vial to 1
— Sequence Type to Injector
— Repeat Sequence to no repeat
6 Start the MSD method from either the ChemStation or the MSD Local
Control Panel.
7 After the "Waiting for remote start ..." status box appears, start the GC
method by:
— Pressing the Start button on the GC Keypad, or
— Pressing the START key (F8) in the Status screen of the GC Control Module, or
— Pressing the START key (F8) in the Status/Automation screen of the GC
Control Module.
To stop data acquisition before the end of the method
• Press the Stop button on the GC Keypad, or
• Press the STOP key (F7) in the Status screen of the GC Control Module.
Pressing either the Stop Seq key (F7) or PAUSE key (F6) in the Status/
Automation screen will not stop data acquisition as these two keys allow the run to
be completed.
When data acquisition is stopped before the end of the run, the MSD ChemStation
will save the data file.
3 Operational Differences
Automatic injection
81