3 Operational Differences
Installing the Column in the GC/MS Interface
8 Loosen the interface nut and push the column an additional 3-5 cm into
the analyzer chamber.
9 Make a clean cut of the column so that now only 3-5 cm protrudes into the
analyzer chamber.
10 Clean the outside of the free end of the column with a lint-free cloth
moistened with methanol.
11 Adjust the column so that it protrudes 1 to 2 mm into the analyzer chamber
past the end of the GC/MS interface, and hand tighten the nut.
Make sure the position of the column does not change as you retighten the nut.
pd¤hm
m›FO"4F˜ pd¤hm˜¤›
— ˜ m›FO"4F˜
| ˜Fm=}
m"d©«F˜ ^"h*F
— ˜ m›FO"4F˜
| ˜Fm=}
s˜›p˜¢˜hh
12 Repeat step 7 to assure column integrity.