Temperature data

Temperature resistance for the connecting cable

The temperature at the cable entries of the device depends on
the design, the measuring medium temperature T
ambient temperature T
amb.
For the electric connection of the device, use only cables with
sufficient temperature resistance in accordance with the
following table.
Devices in integral mount design with dual-compartment
housing
T
amb.
≤ 50 °C (≤ 122 °F)
≤ 60 °C (≤ 140 °F)
≤ 70 °C (≤ 158 °F)
Devices in integral mount design with single-compartment
housing
T
amb.
≤ 50 °C (≤ 122 °F)
≤ 60 °C (≤ 140 °F)
≤ 70 °C (≤ 158 °F)
medium
.
Temperature resistance
≥ 70 °C (≥ 158 °F)
≥ 80 °C (≥ 176 °F)
≥ 90 °C (≥ 194 °F)
Temperature resistance
≥ 75 °C (≥ 167 °F)
≥ 85 °C (≥ 185 °F)
≥ 95 °C (≥ 203 °F)
Sensor in remote mount design
and the
T
amb.
≤ 50 °C (≤ 122 °F)
≤ 60 °C (≤ 140 °F)
≤ 70 °C (≤ 158 °F)
For sensors in remote mount design, the wires in the connection
box must be additionally insulated with the enclosed silicone
hoses starting from ambient temperatures of T
(≥ 140 °F).
Environmental and process conditions for model
FCx4xx...
Ambient temperature T
Measuring medium temperature
T
medium
IP rating / NEMA rating
Optional, with order code 'Ambient temperature range – TA9'
*
Temperature resistance
amb.
13
≥ 105 °C (≥ 221 °F)
≥ 110 °C (≥ 230 °F)
≥ 120 °C (≥ 248 °F)
≥ 60 °C
amb.
-20 to 70 °C
(-4 to 158 °F)
-40 to 70 °C*
(-40 to 158 °F)*
-40 to 205 °C
(-40 to 400 °F)
IP 65, IP 67 /
NEMA 4X,Type 4X