Digital-Analog Board Communication Error Alarm; Calculation Error Alarm; Calibration Un-Normalized Error Alarm; Stream Sequence Error Alarm - ABB NGC8200 User Manual

Gas chromatograph
Table of Contents

Advertisement

11.2.6

Digital-Analog Board Communication Error Alarm

This alarm indicates a communication error between the digital board and the analytical processor
board. Verify the cable connectors are firmly and correctly connected to both the digital and analytical
processor boards.
To troubleshoot digital-analog board communication error:
1.
In the alarm log, check the frequency of the error. If multiple errors exist, place the device in
hold and then launch a cycle.
2.
If the alarms continue to register, perform a warm start.
3.
When the device completes the startup diagnostics without error, place the device in run.
4.
Following 2-3 cycles, verify that no new alarms are registering.
If alarms continue to register, call technical support.
11.2.7

Calculation Error Alarm

This alarm indicates the AGA-8 compressibility calculation is not functioning properly. Typically, this
error would be caused by a gas sample being out of specification for AGA-8; but it could indicate that
the component's peak has shifted.
To troubleshoot calculation error:
1.
Calibrate the analyzer (see section 7), ensuring that the next mode is set to hold.
2.
When the unit enters hold, select Peak Find.
3.
Verify that the peaks are correctly labeled and integrated. If the peaks are not correctly labeled
and integrated, continue to the next step. Otherwise, skip to step 5.
4.
In the Peak Find screen, select Run Manual PF. When the cycle is complete, repeat step 3.
5.
Under Next Mode, select Run.
6.
Allow the unit to run a minimum of an hour and then perform a calibration.
11.2.8

Calibration Un-normalized Error Alarm

This alarm indicates a change to the un-normalized total of sufficient percentage to activate the
alarm. This alarm will discontinue a scheduled calibration and will need to be disabled prior to
calibrating the unit.
To troubleshoot calibration un-normalized error:
1.
On the Analyzer Operation screen, click Hold under Next Mode. When the unit completes the
current cycle and enters hold, continue to the next step.
2.
Compare the calibration blend concentrations to the calibration blend concentrations listed on
the Calibration Setup screen. If an error exists, make corrections and send the setup when
complete.
3.
Under Stream Setup > Alarm Definitions, locate the calibration un-normalized error alarm
and set alarm enable to No. Send change. Repeat for any additional streams with this alarm.
4.
Calibrate the analyzer (see section 7), ensuring that the Next Mode is set to Hold.
5.
When the unit enters hold, select Peak Find.
6.
Verify that the peaks are correctly labeled and integrated. If the peaks are correctly labeled and
integrated, return the unit to operation.
7.
Select Peak Find from the Analyzer Operation screen. Select Run Manual PF.
8.
When the unit enters hold, verify that the peaks are correctly labeled and integrated. If the
peaks are correctly labeled and integrated, return the unit to operation; otherwise, contact
technical support.
9.
Reset the Alarm Enable to Yes. Verify that the alarm threshold is a valid configuration.
Typically, the un-normalized total should be within 50% (between 99.5 and 100.5).
10. Return the unit to regular operation.
11.2.9

Stream Sequence Error Alarm

This alarm indicates a synchronization problem following a manual data post process in factory mode.
To troubleshoot stream sequence error:
1.
On the Analyzer Operation screen, click Hold under Next Mode. When the unit completes the
current cycle and enters hold, continue to the next step.
2.
Perform a warm start (see section 10.9.1).

11.2.10 Calibration CV Percent Error Alarm

This alarm is activated when there is a sufficient change to the CV percent. This alarm will discontinue
a scheduled calibration and will need to be disabled prior to calibrating the unit.
1 2 8 | N G C8 20 0 A N D PG C 1 00 0 US ER MA N UA L | 21 0 15 10 MNA H

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pgc1000

Table of Contents