ABB LGR-ICOS™ GLA232 Series Methane/Ethane Analyzer User Manual
5
Initialize and Run the Analyzer
You may start up the analyzer by doing one of the following:
Use the MobileGuard™ software – refer to the MobileGuard™ System Manual.
Start up the analyzer manually, without the MobileGuard™ software, as described
below.
The analyzer may be controlled by the external software provided with the ABB Ability Mobile
Gas Leak Detection System (MobileGuard™). However, the analyzer can be operated without
this software, as described in this section.
Access Analyzer's Local Operating System
For manual start-up, access the operating system running on the analyzer's internal computer.
This section describes two possible ways to do this: via a remote VNC connection or by a local
connection using a monitor, keyboard, and mouse connected to the analyzer.
Remote Connection
If the analyzer is connected to a LAN, and the IP address is known, access the analyzer's
operating system using a VNC connection (password: lgrvnc).
Local Connection
Plug a VGA compatible monitor into the VGA port on the rear panel of the analyzer, and plug a
USB compatible mouse and keyboard into either the front or rear USB ports on the analyzer.
Analyzer Manual Startup
Once the local operating system has been accessed, perform the following to manually start the
analyzer:
1. Open a terminal window on the desktop: Right-click on the desktop and select Open
Terminal Here.
2. In the terminal window, enter the following commands:
a. cd /usr/icos/
b. ./icos.sh
The ICOS software starts, and the internal pump engages.