Automatic Feed; Directional Fixture For Specimen Clamps - Leica SM2010 R Instructions For Use Manual

Sliding microtome
Table of Contents

Advertisement

5
Operation

5.1.4 Automatic feed

The position of the adjusting knob
where the automatic feed takes place. It should take place immediately in front of the specimen.
Caution
The automatic feed is only effective up to a section thickness of 30 μm. All values set beyond that are
not defined.
• To adjust the automatic feed, move the blade/knife
(→ Fig. 27-3)
and lock it in this position using the locking knob
• Loosen the adjusting button
in this position.
• For exact work with the automatic feed, the knife sledge must be moved past the position of the
adjusting knob.
Note
If the adjusting button

5.1.5 Directional fixture for specimen clamps

Note
All object clamps available as optional accessories can be inserted into the quick clamping device
(→ Fig. 28-5)
of the directional specimen holder fixture, all object clamps available as optional
accessories can be used.
The object orientation allows for simple position correction of the specimen surface when the specimen
is clamped into place.
32
(→ Fig. 27-1)
(→ Fig. 27-1)
and push it back until you can feel resistance. Retighten it
3
2
Fig. 27
(→ Fig. 27-1)
is locked in the frontmost (left) position, there is no feed motion.
determines the point of the knife sledge movement
(→ Fig. 27-2)
until it is just in front of the specimen
(→ Fig.
4
1
27-4).
Version 1.8, Revision I