4.
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement.
The factor that was set last is displayed.
01.09.2017 08:00
6.
Using <> <>, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with <ENTER>.
The meter switches to the measured value display.
The status display [Factor] is active.

8.3.6 Calibration data

The calibration record is automatically transmitted to the interface
after calibrating.
Dissolved oxygen
77