Ion concentration
68
9.
Immerse the electrode in the sample that was supplemented with blank
value solution.
10.
Using <> <> and <ENTER>, enter the values for the volume of the
sample (Sample volume), the volume of the ISA/TISAB solution (ISA/
TISAB vol.), the volume of the blank value solution (BV volume) and the
concentration of the blank value solution (BV conc.).
11.
Select Continue with <> <> and start the measurement with
<ENTER>.
When the measurement is finished an entry window appears.
Blank value addition
Add standard!
i
Std. volume
Std. conc.
Continue
01.09.2017 08:00
12.
Add the standard solution to the sample.
13.
Using <> <> and <ENTER>, enter the values for the volume of the
standard solution (Std. volume) and concentration of the standard solu-
tion (Std. conc.) .
14.
Select Continue with <> <> and start the measurement with
<ENTER>.
The measurement result is displayed when the measurement is com-
pleted.
01.09.2017 08:00
15.
If necessary, start measuring further samples with <ENTER>.
Repeat steps 6 - 12 for all samples.
16.
Terminate the measuring method with <M>.
A safety query appears.
17.
Select yes with <> <>.
18.
Using <ENTER>, confirm yes.
Measurement with the selected method is completed.
100.0 ml
1.0 mg/l
Blank value addition
®
inoLab
Multi 9620 IDS
ba77173e04
12/2017