wtw inoLab Multi 9620 IDS Operating Manual page 52

Digital meter for ids sensors
Table of Contents

Advertisement

Ion concentration
Continuing with
three- to seven-point
calibration
52
15.
When calibrating without temperature sensor:
 Measure the temperature of the standard solution using a thermom-
eter.
 Use <F4>/[ ⇅ ] to select the setting of the temperature.
 Use <> <> to set the temperature.
 Use <F4>/[ ⇅ ] to select the setting of the concentration.
16.
Set the concentration of the standard solution with <> <> and
press <ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
01.09.2017 08:00
17.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with <ENTER>.
The calibration display for the next standard solution appears.
01.09.2017 08:00
18.
Press <ENTER> to continue with three-point calibration.
or
Finish the calibration procedure as a two-point calibration with <M>.
The new calibration values are displayed.
Repeat the steps 12 to 17 in the same way with the third and further standard
solutions as necessary. The new calibration values are displayed after the last
calibration step was completed.
Based on the calibration data, the calibration curve is determined
in sections, according to the Nernst equation modified by Nikolski.
Standard
®
inoLab
Multi 9620 IDS
ba77173e04
12/2017