wtw inoLab Multi 9620 IDS Operating Manual page 126

Digital meter for ids sensors
Table of Contents

Advertisement

What to do if...
No stable measured
value
Obviously incorrect
measured values
126
Cause
IDS pH sensor:
– Junction contaminated
– Membrane contaminated
Test sample:
– pH value not stable
– Temperature not stable
IDS pH sensor + test sample:
– Conductivity too low
– Temperature too high
– Organic liquids
Cause
IDS pH sensor:
– IDS pH sensor unsuitable
– Temperature difference between
buffer and test sample too great
– Measurement procedure not suit-
able
®
inoLab
Multi 9620 IDS
Remedy
– Clean the junction
– Clean membrane
– Measure with air excluded if
necessary
– Adjust temperature if necessary
– Use suitable IDS pH sensor
– Use suitable IDS pH sensor
– Use suitable IDS pH sensor
Remedy
– Use suitable IDS sensor
– Adjust temperature of buffer or
sample solutions
– Follow special procedure
ba77173e04
12/2017