wtw inoLab Multi 9620 IDS Operating Manual page 99

Digital meter for ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9620 IDS
ba77173e04
12/2017
Menu item
Man. temperature
Alternative tempera-
ture
ISE setup/
AutoRead criterion
ISE setup/
Ion type
ISE setup/
Unit
ISE setup/
Valency
ISE setup/
Molar mass
ISE setup/
Density
Method
Start method
Possible setting
-25 ... +25 ... +130 °C
on
off
low
medium
high
Ag, Br, Ca, Cd, Cl,
CN, Cu, F, I, K, Na,
NO3, Pb, S, NH3,
NH4, CO2, ION
* Measuring with the NH 500 electrode:
The NH4 setting is not suitable for the gas-sensitive
electrode NH 500.
Select the following settings:
Ion type "ION", Valency "-1".
mg/l
µmol/l
mg/kg
ppm
%
-8 ... +8
1 ... 300 g/mol
0.001 ... 9.999 g/ml or
kg/l
Standard addition
Standard subtraction
Sample addition
Sample subtraction
Blank value addition
Explanation
Entry of the manually deter-
mined temperature. For mea-
surements without
temperature sensor only.
Takes over the temperature
value from an IDS sensor.
This menu item is only avail-
able if an IDS adapter and an
IDS sensor with integrated
temperature sensor are con-
nected.
Selection of the AutoRead
criteria (see section 7.1.1
M
EASURING THE ION CONCEN
, page 47).
TRATION
Selection of the ion type to be
measured.
An ion that is not included in
the list can be measured with
the setting, ION.
Selection, with which unit the
measurement result and cali-
bration standards should be
displayed.
Set the valence (Valency)
and molar weight (Molar
mass) for the ion
(only for ISE setup/Ion type =
ION)
Adjustable density of the test
sample (only with Unit: mg/
kg, ppm, %)
Selection of the available
measuring methods.
Start measurement with the
selected method.
Settings
-
99