wtw inoLab Multi 9620 IDS Operating Manual page 110

Digital meter for ids sensors
Table of Contents

Advertisement

Settings
Conductivity
110
The following settings for conductivity measurements are reset to the default
settings with the Reset function:
Setting
Calibration interval
Measured parameter
Cell constant (c)
Temperature compensation
Reference temperature
Temperature coefficient (TC) of
the linear temperature compensa-
tion
TDS multiplier
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <ENTER>
key.
inoLab
Default settings
150 d
ϰ
Depending on the connected measur-
ing cell:
-1
0.475 cm
(calibrated)
-1
0.475 cm
(set)
-1
0.100 cm
nLF
25 °C
2.000 %/K
1.00
ba77173e04
®
Multi 9620 IDS
12/2017