Data storage
Starting the auto-
matic storing func-
tion
Terminating the
automatic memory
function prema-
turely
114
To start the automatic storing function, select Continue with <><> and con-
firm with <ENTER>. The meter switches to the measured value display.
0d03h00min
01.09.2017 08:00
1 Remaining storing duration
2 Graphical display of the storing duration
The active automatic data storing function can be recognized by the progress
bar in the status line. The progress bar indicates the remaining storage dura-
tion.
If the automatic data storage function is activated, only the follow-
ing keys are active: <M><><><STO_> and <On/Off>. The
other keys and the automatic switch-off function are deactivated.
Proceed as follows to switch off the automatic data storing function before the
adjusted storing duration has expired:
1.
Press the <STO_> key.
The following window appears.
Warning
Stop automatic storage?
yes
no
01.09.2017 08:00
2.
Using <><>, select yes and confirm with <ENTER>.
The meter switches to the measured value display.
The automatic data storing function is terminated.
®
inoLab
Multi 9620 IDS
1
2
ba77173e04
12/2017