Automatic data storage
ID number
Interval
Duration
Continue
0d03h00min
0
1 Specified entire storing duration
2 Max. available storing duration
3 Graphical display of the memory usage
You can configure the automatic data storing function with the following
settings:
Menu item
Possible setting
ID number
1 ... 10000
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min, 10 min,
15 min, 30 min,
60 min
Duration
1 min ... x min
1
30 s
180 min
1d17h33min
01.09.2017 08:00
Explanation
ID number for the dataset series.
Storing interval.
The lower limit of the storing interval
can be restricted by the number of free
memory locations.
The upper limit is restricted by the stor-
ing duration.
Storing duration.
Specifies after which time the auto-
matic data storing should be termi-
nated.
The lower limit of the storage duration
is restricted by the storage interval.
The upper limit is restricted by the
number of free memory locations.