Ion concentration
Calibration evalua-
tion
54
Menu item
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to USB
flash drive or printer
Calibration /
Calibration data stor-
age / Output to
RS232/USB
After calibrating, the meter automatically evaluates the calibration.
Display
Calibration record
+++
Error
Error
Error elimination (see section
15.2 ISE, page 127)
Setting/
Explanation
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
records with <><>.
 Using <PRT>, output the dis-
played calibration record to the
interface USB-B (USB Device,
e.g. PC) or USB-A (USB Host,
e.g. USB printer).
 Using <PRT_>, output all calibra-
tion records to the interface USB-
B (USB Device, e.g. PC) or USB-
A (USB Host, e.g. USB printer).
 Quit the display with <ESC>.
 Switch directly to the measured
value display with <M>.
-
Outputs the stored calibration data to
the USB-A interface (USB Host, e.g.
USB memory device or USB printer).
-
Outputs the stored calibration data to
the USB-B interface (USB Device,
e.g. PC)
Magnitude of the slope [mV]
30.0 ... 90.0 *
or 15.0 ... 45.0 **
< 30 or > 90 *
or < 15 or > 45 **
*
Monovalent ions
** Bivalent ions
®
inoLab
Multi 9620 IDS
ba77173e04
12/2017