60
9.
Immerse the electrode in the sample.
10.
Using <> <> and <ENTER>, select the values for the volume of the
sample (Sample volume) and the volume of the ISA/TISAB solution
(ISA/TISAB vol.).
11.
Select Continue with <> <> and start the measurement with
<ENTER>.
When the measurement is finished an entry window appears.
Standard subtraction
Add standard!
i
Ion type
Std. volume
Std. conc.
Continue
01.09.2017 08:00
12.
Add the standard (complexing agent or precipitating agent) to the sam-
ple.
13.
Using <> <> and <ENTER>, enter the values for the volume of the
standard solution (Std. volume) and concentration of the standard solu-
tion (Std. conc.) .
14.
Using <> <> and <ENTER>, set the ion type for the ion in the stan-
dard (Ion type) if necessary.
On selection of the ion type that can be defined (ION):
Using <> <> and <ENTER>, set the valence (Valency) and molar
weight (Molar mass) for the ion in the standard solution.
15.
Select Continue with <> <> and start the measurement with
<ENTER>.
The measurement result is displayed when the measurement is com-
pleted.
01.09.2017 08:00
16.
If necessary, start measuring further samples with <ENTER>.
Repeat steps 6 - 15 for all samples.