wtw inoLab Multi 9620 IDS Operating Manual page 90

Digital meter for ids sensors
Table of Contents

Advertisement

Turbidity measurement (VisoTurb
Continuing with two-
point calibration
90
®
900-P)
6.
Thoroughly rinse the turbidity sensor with distilled water and dry it with
a lint-free cloth.
7.
Immerse the turbidity sensor in the test sample at an angle.
8.
Position the turbidity sensor in the measuring vessel.
9.
Use <> <> and <><> to set the concentration of the standard
solution for each digit and confirm with <ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
10.
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
Standard
Cal
124.0 FNU
15.09.2014 08:00
11.
Thoroughly rinse the turbidity sensor with distilled water and dry it with
a lint-free cloth.
12.
Immerse the turbidity sensor in the test sample at an angle.
13.
Position the turbidity sensor in the measuring vessel.
14.
Use <> <> and <><> to set the concentration of the standard
solution for each digit and confirm with <ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
15.
Wait for the end of the AutoRead measurement.
The calibration display for the next standard solution appears.
Standard
Cal
1010.0 FNU
15.09.2014 08:00
2
3
®
inoLab
Multi 9620 IDS
ba77173e04
12/2017