wtw inoLab Multi 9620 IDS Operating Manual page 51

Digital meter for ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9620 IDS
Continuing with two-
point calibration
ba77173e04
12/2017
8.
Thoroughly rinse the electrode with distilled water.
9.
Immerse the electrode in standard solution 1.
10.
When calibrating without temperature sensor:
 Measure the temperature of the standard solution using a thermom-
eter.
 Use <F4>/[ ⇅ ] to select the setting of the temperature.
 Use <> <> to set the temperature.
 Use <F4>/[ ⇅ ] to select the setting of the concentration.
11.
Set the concentration of the standard solution with <> <> and press
<ENTER>.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
01.09.2017 08:00
12.
Wait for the end of the AutoRead measurement or accept the calibration
value with <ENTER>.
The calibration display for the next standard solution appears.
01.09.2017 08:00
13.
Thoroughly rinse the electrode with distilled water.
14.
Immerse the electrode in standard solution 2.
Standard
Ion concentration
51

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Inolab multi 9630 ids

Table of Contents