wtw inoLab Multi 9620 IDS Operating Manual page 34

Digital meter for ids sensors
Table of Contents

Advertisement

pH value
Continuing with two-
point calibration
34
9.
Set the nominal buffer value for the measured temperature with
<><>.
10.
Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
11.
If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS-
pH sensor.
12.
Thoroughly rinse the sensor with deionized water.
13.
Immerse the pH sensor in buffer solution 2.
14.
When measuring without temperature sensor:
Enter the temperature of the buffer with <><>.
15.
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
16.
Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>.
The pH value of the buffer solution is displayed.
ConCal
01.09.2017 08:00
17.
Set the nominal buffer value for the measured temperature with
<><>.
18.
Accept the calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
19.
If necessary, finish the calibration procedure as a two-point calibration
with <M>.
The calibration record is displayed.
®
inoLab
Multi 9620 IDS
ba77173e04
12/2017