wtw inoLab Multi 9620 IDS Operating Manual page 37

Digital meter for ids sensors
Table of Contents

Advertisement

®
inoLab
Multi 9620 IDS
Calibration evalua-
tion
ba77173e04
12/2017
Menu item
Calibration /
Calibration data stor-
age / Display
Calibration /
Calibration data stor-
age / Output to USB
flash drive or printer
Calibration /
Calibration data stor-
age / Output to
RS232/USB
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
Display
Calibration
record
+++
++
+
Setting/
Explanation
function
-
Displays the calibration records.
Further options:
 Scroll through the calibration
records with <><>.
 Using <PRT>, output the dis-
played calibration record to the
interface USB-B (USB Device,
e.g. PC) or USB-A (USB Host,
e.g. USB printer).
 Using <PRT_>, output all calibra-
tion records to the interface USB-
B (USB Device, e.g. PC) or USB-
A (USB Host, e.g. USB printer).
 Quit the display with <ESC>.
 Switch directly to the measured
value display with <M>.
-
Outputs the stored calibration data to
the USB-A interface (USB Host, e.g.
USB memory device/USB printer).
-
Outputs the stored calibration data to
the USB-B interface (USB Device,
e.g. PC)
Zero point
[mV]
-15 ... +15
-20 ... <-15
or
>+15 ... +20
-25 ... <-20
or
>+20 ... +25
pH value
Slope [mV/pH]
-60.5 ... -58.0
>-58.0 ... -57.0
-61.0 ... <-60.5
or
>-57.0 ... -56.0
37