Plantweb® Alerts - Emerson Rosemount 5400 Series Reference Manual

Superior performance two-wire non-contacting radar level transmitter
Hide thumbs Also See for Rosemount 5400 Series:
Table of Contents

Advertisement

Appendix H: Resource block
November 2014
Parameter
ADVISE_MASK
ADVISE_ACTIVE
ADVISE_ALM
HEALTH_INDEX
PWA_SIMULATE
H.2.1
PlantWeb
The Resource Block will act as a coordinator for PlantWeb alerts. There will be three alarm
parameters (FAILED_ALARM, MAINT_ALARM, and ADVISE_ALARM) which will contain
information regarding some of the device errors which are detected by the transmitter
software. There will be a RECOMMENDED_ACTION parameter which will be used to display the
recommended action text for the highest priority alarm and a HEALTH_INDEX parameters (0 -
100) indicating the overall health of the transmitter. FAILED_ALARM will have the highest
priority followed by MAINT_ALARM and ADVISE_ALARM will be the lowest priority.
FAILED_ALARMS
A failure alarm indicates a failure within a device that will make the device or some part of the
device non-operational. This implies that the device is in need of repair and must be fixed
immediately. There are five parameters associated with FAILED_ALARMS specifically, they are
described below.
FAILED_ENABLED
This parameter contains a list of failures in the device which makes the device non-operational
that will cause an alert to be sent. Below is a list of the failures:
FAILED_MASK
This parameter will mask any of the failed conditions listed in FAILED_ENABLED. A bit on means
that the condition is masked out from alarming and will not be reported.
278
Index
Default value
number
81
0x00000000
(0x00000000)
82
0x00000000
(0x00000000)
83
84
100
85
0x00(Simulation off)
®
alerts
Electronics Failure - Output Board
Internal Communication Failure
Memory Failure - Output Board
Software Incompatibility Error
Electronics Failure - Main Board
Internal Temperature Critical
Sensor Database Error
Software Error
Description
Mask of ADVISE_ALM. Corresponds bit of bit to
ADVISE_ACTIVE. A bit on means that the condition is
masked out from alarming.
Enumerated list of advisory conditions within a device.
Alarm indicating advisory alarms. These conditions do
not have a direct impact on the process or device
integrity.
Parameter representing the overall health of the
device, 100 being perfect and 10 being
non-functioning. The value is based on the active PWA
alarms.
Parameter allowing simulation of PWA alarms.
Reference Manual
00809-0100-4026, Rev HA
Resource Block

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 5402Rosemount 5401

Table of Contents