Reference Manual
00809-0100-4026, Rev HA
Appendix E
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 257
Parameters and descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 258
Supported units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 263
Diagnostics device errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 264
E.1
Overview
This section contains information on the Rosemount 5400 Series Transducer Block (TB).
Descriptions of all Transducer Block parameters, errors, and diagnostics are listed.
E.1.1
Definition
The transducer block contains the actual measurement data, including a level and distance
reading. Channels 1–6 are assigned to these measurements. The transducer block includes
information about sensor type, engineering units, and all parameters needed to configure the
radar gauge.
E.1.2
Channel definitions
Each input has an assigned channel which can be linked to the AI block. The channels for the
Rosemount 5400 Series are the following:
Table E-1. Channel Assignments
Channel name
Level
Ullage
Level Rate
Signal Strength
Volume
Internal Temperature
Level Transducer Block
Level Transducer Block
Channel number Process variable
1
2
3
4
5
6
Appendix E: Level transducer block
RADAR_LEVEL
RADAR_ULLAGE
RADAR_LEVELRATE
RADAR_LEVEL_SIGNAL_STRENGTH
RADAR_VOLUME
RADAR_INTERNAL_TEMPERATURE
November 2014