Emerson Rosemount 5400 Series Reference Manual page 295

Superior performance two-wire non-contacting radar level transmitter
Hide thumbs Also See for Rosemount 5400 Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-4026, Rev HA
Failed/maint/advise/
Alarm type
active event
Software error
Sensor database error
Internal temperature
critical
Electronics failure –
main board
Failed
Software
incompatibility error
Memory failure -
output board
Internal
communication failure
Electronics failure -
output board
Resource Block
Description
The device software has encountered an
error. Reasons may be multiple, including
too low supplied voltage, or an error
being simulated.
The status for all variables is BAD, and the
device is out of service.
The device has found an error in the
configuration database.
Status for all variables is BAD. Device
recovery is possible.
The internal temperature of the device
has reached critical levels and the
integrity of the device electronics has
been compromised. Ambient
temperature should not exceed device
specifications
(-40 °F/-40 °C to 176 °F/+80 °C).
The device is not in service (OOS) and
status for all variables is BAD.
The device has detected a fault with an
electrical component on the main board
electronics module assembly.
The status for all variables is BAD, and the
device is out of service.
Fieldbus software and main firmware
versions are incompatible.
The device is not in service (OOS).
Configuration data has been corrupted or
pending configuration changes has been
lost due to loss of power before storage
could complete.
Default values are loaded into the faulty
block. Potential errors in stored data may
cause unwanted behavior. The device is
not in service (OOS) and status for all
variables is BAD. Device recovery is
possible.
The communication between the main
transmitter board and the fieldbus
electronics board has been lost.
The device holds the last known
measurement values with status BAD.
The device has detected a fault with an
electrical component on the output
board electronics module assembly.
The device is not in service (OOS).
Appendix H: Resource block
Recommended action
1. Make sure that enough voltage is
supplied to the device.
2. Restart the device.
3. Turn off simulation of device failure.
4. If the alert persists, replace the
transmitter head.
1. Load default database to the device
to clear the error.
2. Reconfigure the device.
3. If the alert persists, replace the
transmitter head.
Replace the transmitter head.
Replace the transmitter head.
Replace the transmitter head.
1. Load default database to the device
to clear the error.
2. Download a device configuration.
3. If the error persists, it may indicate a
faulty memory chip. Replace the
transmitter head.
Replace the transmitter head.
Replace the transmitter head.
November 2014
283