Lcd Display Error Messages - Emerson Rosemount 5400 Series Reference Manual

Superior performance two-wire non-contacting radar level transmitter
Hide thumbs Also See for Rosemount 5400 Series:
Table of Contents

Advertisement

Section 6: Operation
November 2014
6.3

LCD display error messages

Figure 6-7. The Rosemount 5400 Series Display Panel Displaying an Error Message
Table 6-2. Error Messages Displayed on the Rosemount 5400 Series Display Panel
Error message
RAM FAIL
FPROM FAIL
HREG FAIL
OMEM FAIL
MWM FAIL
DPLY FAIL
MODEM FAIL
AOUT FAIL
OHW FAIL
ITEMP FAIL
MEAS FAIL
CONFIG FAIL
SW FAIL
For more information on errors, see
126
Description
An error in the gauge data memory (RAM) has been detected during the
startup tests.
NOTE: this resets the gauge automatically.
An error in the gauge program memory (FPROM) has been detected
during the startup tests.
NOTE: this resets the gauge automatically.
An error in the transmitter configuration memory (EEPROM) has been
detected. The error is either a checksum error that can be solved by
loading the default database or a hardware error.
NOTE: the default values are used until the problem is solved.
Other memory failure.
An error in the microwave module.
An error in the LCD display.
Modem hardware failure.
An error in the Analog Out Module.
An unspecified hardware error has been detected.
An error in the internal temperature measurement.
A serious measurement error has been detected.
At least one configuration parameter is outside the allowed range.
NOTE: the default values are used until the problem is solved.
An error has been detected in the transmitter software.
"Error Messages" on page
Error message
154.
Reference Manual
00809-0100-4026, Rev HA
Operation