Emerson Rosemount 5400 Series Reference Manual page 288

Superior performance two-wire non-contacting radar level transmitter
Hide thumbs Also See for Rosemount 5400 Series:
Table of Contents

Advertisement

Appendix H: Resource block
November 2014
Parameter
FB_OPTIONS
DIAG_OPTIONS
MISC_OPTIONS
RB_SFTWR_REV_MAJOR
RB_SFTWR_REV_MINOR
RB_SFTWR_REV_BUILD
RB_SFTWR_REV_ALL
hardware_rev
output_board_sn
FINAL_ASSY_NUM
detailed_status
summary_status
message_date
message_text
SELF_TEST
define_write_lock
save_config_now
save_config_blocks
start_with_defaults
simulate_IO
security_IO
276
Index
Default value
number
45
0x00000000
(0x00000000)
46
0x00000000
(0x00000000)
47
0x00000000
(0x00000000)
48
3
49
0
50
33
51
3-00-033 - Mon Apr 18
16:47:05 2011 by
chadwar
52
6
53
54
0
55
0x00000000
(0x00000000)
56
0x01(No repair needed)
57
01/01/84 00:00:00
58
59
0x01(No self test)
60
0x01(Everything locked)
61
0x01(No save)
62
0
63
0x01(No NV defaults)
64
0x01(Jumper off)
65
0x01(Jumper off)
Description
Indicates which function block licensing options are
enabled.
Indicates which diagnostics licensing options are
enabled.
Indicates which miscellaneous licensing options are
enabled.
Major revision of software that the resource block was
created with.
Minor revision of software that the resource block was
created with.
Build of software that the resource block was created
with.
The string will contain the following fields:
Major rev: 1-3 characters, decimal number 0-255
Minor rev: 1-3 characters, decimal number 0-255
Build rev: 1-5 characters, decimal number 0-255
Time of build: 8 characters, xx:xx:xx, military time
Day of week of build: 3 characters, Sun, Mon,...
Month of build: 3 characters, Jan, Feb.
Day of month of build: 1-2 characters, decimal number
1-31
Year of build: 4 characters, decimal
Builder: 7 characters, login name of builder
Hardware revision of the hardware that has the
resource block in it.
Output board serial number.
The same final assembly number placed on the neck
label.
Indicates the state of the transmitter. See Resource
Block detailed status codes.
An enumerated value of repair analysis.
Date associated with the MESSAGE_TEXT parameter.
Used to indicate changes made by the user to the
device's installation, configuration,
or calibration.
Instructs resource block to perform self-test. Tests are
device specific.
Allows the operator to select how WRITE_LOCK
behaves. The initial value is "lock everything". If the
value is set to "lock only physical device", then the
resource and transducer blocks of the device will be
locked but changes to function blocks will be allowed.
Allows the user to optionally save all non-volatile
information immediately.
Number of EEPROM blocks that have been modified
since last burn. This value will count down to zero
when the configuration is saved.
0 = Uninitialized
1 = do not power-up with NV defaults
2 = power-up with default node address
3 = power-up with default pd_tag and node address
4 = power-up with default data for the entire
communications stack (no application data)
Status of simulate switch.
Status of security switch.
Reference Manual
00809-0100-4026, Rev HA
Resource Block

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 5402Rosemount 5401

Table of Contents