Section 5: Basic Configuration/Start-up
November 2014
98
Restart the transmitter
15. Restart the transmitter to activate all of the
configuration changes. It may take up to 60
seconds after the restart button is pressed
before measurement values are updated.
Verify Level
16. Run the Verify Level tool to match the product
level reported by the device to a reference
measurement (measured by using for example
hand gauging). In any difference, the Calibration
Distance parameter will be adjusted.
A minor adjustment using Calibration Distance is
normal. There may, for example be a deviation
between the actual tank height and the
configured value.
Configuration backup
17. When configuration is complete, the
configuration should be saved to a backup file.
This information is useful for:
installing another Rosemount 5400 Series
transmitter in a similar tank, since the file
can be directly uploaded to a new device.
restoring the configuration, if configuration
data is lost or accidentally modified, making
the device inoperable.
Use Archive Device to create a backup file and
save additional information, such as an echo
curve movie, for future reference to a .zip
file.
Reference Manual
00809-0100-4026, Rev HA
Basic Configuration/Start-up