13. User calibration

If several measuring runs are necessary to balance a wheel
1
because balance weight size and position have to be changed
repeatedly, it may be necessary to readjust the machine.
For user calibration a calibration weight is supplied with the
machine (ref. no. 6418 416 - kept on the right at the back of
the weight tray).
Important:
Readjustment must be carried out using the clamping
adapter supplied with the machine.
Readjustment (Code C14)
• Make sure no wheel or other clamping means is
clamped on the machine.
• Press and hold the CODE key (Fig.130, item 1)
The basic reading C -- is read out (Fig. 131).
• Release
Reading 1. and START are read out (Fig. 132).
• Close the optional wheel guard, and press the START
On completion of the first run reading 2. and START are read
out (Fig. 133).
• Screw the calibration weight into the threaded hole of
• Press the START key.
On completion of the processing operation a three-tone signal
is given and readjustment is completed.
• Remove the calibration weight from the wheel adapter
Operation manual geodyna 2600 EEWB703D
t he CODE key.
key to carry out a readjustment run.
the wheel adapter (Fig. 134).
and put it back in its designated place.
User calibration