Changing modes of operation
Input: Code/state
-
to 6
-
to 7
71
-
to 7
-
to 60
72
-
to 8
+
Present limit value
-
+
to 5.50
73
36
Display: Code/state
C 6
C 6
7
C 7
C 7
60
C 8
C 8
3.50
C 8
5.50

Code C6

Fig. 71 Number of revolutions per measuring run
5 to 25 revolutions possible,
factory-set to 10*
Example: change to 7 revolutions per measuring run
Note
Reducing the number of measurement revolutions will reduce
the accuracy of measurement. The selected mode of opera-
tion can be transferred to the permanent memory.

Code C7

Fig. 72 Volume of audible signals Scale of
volume 0 to 100 (low - high), factory-
adjusted to 50*
Example: set volume to 60
The volume is not changed before the CODE key is pressed
for quitting the mode.
The selected mode of operation can be transferred to the
permanent memory.

Code C8

Fig. 73 Selecting the limit (threshold) value for suppression
of minor unbalance readings in grammes, or ounces.
The unit (gms or oz) depends on the settings made
under C3.
Grammes:
Range 3.50 to 20.0 g
Factory-adjusted to 3.5 g
Read out limit, e. g. 3.50 g
Select another limit, e. g. 5.50 g
Ounces:
Range 0.25 to 2.00 oz
Factory-adjusted to 0.25 oz
Read out limit, e. g. 0.25 oz
Select another limit, e. g. 0.50 oz
The selected mode of operation can be transferred to the
permanent memory.
* = Factory adjusted mode
Operation manual geodyna 2600 EEWB703D