GN610, GN611 Isolation and type testing
3.2
Channel to channel isolation test
For the channel to channel test, both the DC and AC tests below (see
Figure 3.3 and Figure 3.4) are used with DC voltage √2 higher than the AC
voltage. The test value meets the requirements for 600 V CAT II REINFORCED,
the value for 2000 V RMS BASIC is lower and therefore also covered with this
test. Tests are conducted for one minute, see IEC61010-1 for details.
Figure 3.3: DC Type test Channel to Channel
HBM public
11