To Change the Septum on the Split/Splitless Inlet

Materials needed
• Gloves, clean, lint-free
• Replacement septum
• Wrench, hex for changing septum
• 0- or 00-grade steel wool (optional)
• Tweezers
• Wrench, capillary inlet (optional)
Procedure
1 Set the inlet pressure to 0 and wait for the GC to become ready.
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is
WA R N I N G
hot, wear heat-resistant gloves to protect your hands.
2 Remove the septum retainer nut or Merlin cap.
3 Use tweezers to remove the septum or Merlin Microseal from the retainer
5975T LTM GC/MSD Operation Manual
Large (8650-0030)
Small (8650-0029)
Septum retainer nut for headspace (18740-60830)
Septum retainer nut (18740-60835)
11-mm septum, high-temperature, low-bleed, 50/pk (5183-4757)
11-mm septum, prepierced, long life, 50/pk (5183-4761)
Merlin Microseal septum (high-pressure) (5182-3444)
Merlin Microseal septum (30 psi) (5181-8815)
nut. Do not gouge or scratch the interior of the septum head.