Operating Instructions — Analyze and Process Data
2. Click Explore > Extract Ions > Use Base Peak Masses.
An XIC of the specified selection opens below the spectrum pane. The experiment
information at the top of the pane shows the mass range and the maximum intensity
in counts per second.
Extract Ions by Selecting Masses
1. Open a spectrum or chromatogram.
2. Click Explore > Extract Ions > Use Dialog.
Figure 10-3 Extract Ions Dialog
3. Type the values for each XIC to be created. If a stop value is not typed, then the
range is defined by the start value.
In the Start field, type the start value (lower value) for the mass range.
In the Stop field, type the stop value (higher value) for the mass range.
4. Click OK.
An XIC of the selection opens below the chromatogram pane. The experiment
information at the top of the pane includes the masses and the maximum intensity in
counts per second.

Generate BPCs

BPCs can be generated only from single period, single experiment data. For more information
about using the available icons, refer to
1. Select an area within a TIC.
4500 Series of Instruments
84 of 138
Table
10-8.
System User Guide
RUO-IDV-05-0264-A