Appendix A Parameters For 4500 Series Instruments - AB Sciex QTRAP 4500 LC/MS/MS system System User's Manual

Qtrap 4500 series
Table of Contents

Advertisement

Parameters for 4500 Series Instruments
The following table contains generic parameters for the 4500 series of instruments. The first
number under each scan type is the preset value; the range of numbers is the accessible range
for each parameter.
Table A-1
4500 Series Instrument Parameters
Access ID
Positive ion mode
Q1
(6)
CUR
20
10 to 55
(1)(2)(3)(4)
CUR
10
(5)
10 to 55
(a)
CAD
0
Fixed
(b)
CAD
0
Fixed
(1)(2)
IS
5500
0 to 5500
(5)
IS
1500
Ion Transfer
0 to 2500
Voltage
(6)
IS
1000
0 to 4000
(3)(4)
NC
3
0 to 5
(2)(3)(4)(5)
TEM
0
0 to 750
(a)
DP
200
0 to 300
(b)
DP
130
0 to 300
(1) Turbo V™ ion source (2) TurboIonSpray
source (1=TIS, and 2=HN) (5) PhotoSpray
(a) AB SCIEX QTRAP
System User Guide
RUO-IDV-05-0264-A
Q3
MS/MS
20
20
10 to 55
10 to 55
10
10
10 to 55
10 to 55
6
Medium
Fixed
Low,
Medium,
High
5
9
Fixed
0 to 12
5500
5500
0 to 5500
0 to 5500
1500
1500
0 to 2500
0 to 2500
1000
1000
0 to 4000
0 to 4000
3
3
0 to 5
0 to 5
0
0
0 to 750
0 to 750
100
100
0 to 300
0 to 300
130
120
0 to 300
0 to 300
®
®
ion source (6) NanoSpray
®
4500 system (b) AB SCIEX Triple Quad™ 4500 system
Negative ion mode
Q1
Q3
20
20
10 to 55
10 to 55
10
10
10 to 55
10 to 55
0
6
Fixed
Fixed
0
5
Fixed
Fixed
–4500
–4500
–4500 to 0
–4500 to 0
–1500
–1500
–2500 to 0
–2500 to 0
–1000
–1000
–4000 to 0
–4000 to 0
–3
–3
–5 to 0
–5 to 0
0
0
0 to 750
0 to 750
–100
–100
–300 to 0
–300 to 0
–60
–60
–300 to 0
–300 to 0
probe (3) APCI probe (4) DuoSpray™ ion
A
MS/MS
20
10 to 55
10
10 to 55
Medium
Low,
Medium,
High
9
0 to 12
–4500
–4500 to 0
–1500
–2500 to 0
–1000
–4000 to 0
–3
–5 to 0
0
0 to 750
–100
–300 to 0
–150
–300 to 0
®
ion source
4500 Series of Instruments
131 of 138